TIGHTNESS TEST FAILED PADA VENTILATOR RAPHAEL

SERING TERJADI KETIKA KITA MELAKUKAN KALIBRASI TERNYATA FAILED. BERIKUT AKAN SAYA BAGIKAN INFORMASI DAN TIPS MENGENAI MENGATASI TIGHTNESS TEST FAILED PADA VENTILATOR RAPHAEL.

GAMBAR TIGHTNESS TEST FAILED

YANG PERLU KITA CEK ADALAH :
  1. SAMBUNGAN-SAMBUNGAN TUBING ( MUNGKIN ADA YANG BOCOR,LUBANG ATAU LEPAS )
  2. CEK KARET MEMBRANE WARNA MERAH DI EKSPIRASI ( LIHAT GAMBAR DIBAWAH INI ) JIKA KOTOR DIBERSIHKAN, JIKA DITERAWANG BERLUBANG SEGERA DIGANTI.
LOKASI KARET MEMBRANE

DIPUTAR DAN DILEPAS

NAMPAK DARI ATAS

MEMBRANE JIKA DIBALIK. JIKA KOTOR SEPERTI GAMBAR DIATAS, BISA DIBERSIHKAN DENGAN ALKOHOL SWAB ATAUPUN KASA STERIL DIBASAHI WATER STERIL

MEMBRANE SETELAH DIBERSIHKAN

3. BISA JUGA KEBOCORAN PADA WATER TRAP ( YANG SERING DIPUTAR UNTUK DIBUANG AIRNYA). BISA DIGANTI DENGAN WATER TRAP YANG BARU ATAU DIPASANGKAN DENGAN WATER TRAP LAINNYA YG BELUM TERPAKAI
WATER TRAP
4. BISA JUGA TIGHTNESS TEST FAILED DARI KEBOCORAN DI FLOW SENSOR BISA PECAH ATAPUN SAMBUNGAN

CONTOH SAMBUNGAN YG SALAH

0 komentar:

Posting Komentar