SERING TERJADI KETIKA KITA MELAKUKAN KALIBRASI TERNYATA FAILED. BERIKUT AKAN SAYA BAGIKAN INFORMASI DAN TIPS MENGENAI MENGATASI TIGHTNESS TEST FAILED PADA VENTILATOR RAPHAEL.
 |
GAMBAR TIGHTNESS TEST FAILED
YANG PERLU KITA CEK ADALAH :
SAMBUNGAN-SAMBUNGAN TUBING ( MUNGKIN ADA YANG BOCOR,LUBANG ATAU LEPAS )
- CEK KARET MEMBRANE WARNA MERAH DI EKSPIRASI ( LIHAT GAMBAR DIBAWAH INI ) JIKA KOTOR DIBERSIHKAN, JIKA DITERAWANG BERLUBANG SEGERA DIGANTI.
|
 |
LOKASI KARET MEMBRANE |
 |
DIPUTAR DAN DILEPAS |
 |
NAMPAK DARI ATAS |
 |
MEMBRANE JIKA DIBALIK. JIKA KOTOR SEPERTI GAMBAR DIATAS, BISA DIBERSIHKAN DENGAN ALKOHOL SWAB ATAUPUN KASA STERIL DIBASAHI WATER STERIL |
 |
MEMBRANE SETELAH DIBERSIHKAN |
3. BISA JUGA KEBOCORAN PADA WATER TRAP ( YANG SERING DIPUTAR UNTUK DIBUANG AIRNYA). BISA DIGANTI DENGAN WATER TRAP YANG BARU ATAU DIPASANGKAN DENGAN WATER TRAP LAINNYA YG BELUM TERPAKAI
 |
WATER TRAP |
4. BISA JUGA TIGHTNESS TEST FAILED DARI KEBOCORAN DI FLOW SENSOR BISA PECAH ATAPUN SAMBUNGAN
 |
CONTOH SAMBUNGAN YG SALAH |