SERING TERJADI KETIKA KITA MELAKUKAN KALIBRASI TERNYATA FAILED. BERIKUT AKAN SAYA BAGIKAN INFORMASI DAN TIPS MENGENAI MENGATASI TIGHTNESS TEST FAILED PADA VENTILATOR RAPHAEL.
|
GAMBAR TIGHTNESS TEST FAILED
YANG PERLU KITA CEK ADALAH :
SAMBUNGAN-SAMBUNGAN TUBING ( MUNGKIN ADA YANG BOCOR,LUBANG ATAU LEPAS )
- CEK KARET MEMBRANE WARNA MERAH DI EKSPIRASI ( LIHAT GAMBAR DIBAWAH INI ) JIKA KOTOR DIBERSIHKAN, JIKA DITERAWANG BERLUBANG SEGERA DIGANTI.
|
|
LOKASI KARET MEMBRANE |
|
DIPUTAR DAN DILEPAS |
|
NAMPAK DARI ATAS |
|
MEMBRANE JIKA DIBALIK. JIKA KOTOR SEPERTI GAMBAR DIATAS, BISA DIBERSIHKAN DENGAN ALKOHOL SWAB ATAUPUN KASA STERIL DIBASAHI WATER STERIL |
|
MEMBRANE SETELAH DIBERSIHKAN |
3. BISA JUGA KEBOCORAN PADA WATER TRAP ( YANG SERING DIPUTAR UNTUK DIBUANG AIRNYA). BISA DIGANTI DENGAN WATER TRAP YANG BARU ATAU DIPASANGKAN DENGAN WATER TRAP LAINNYA YG BELUM TERPAKAI
|
WATER TRAP |
4. BISA JUGA TIGHTNESS TEST FAILED DARI KEBOCORAN DI FLOW SENSOR BISA PECAH ATAPUN SAMBUNGAN
|
CONTOH SAMBUNGAN YG SALAH |
0 komentar:
Posting Komentar
In Here